PART
|
96T
|
48T
|
STORAGE
|
|
Pre-Coated Plate
|
8×12
|
8×6
|
4 ℃ / -20 ℃
|
|
Pre-Mixed Standards
|
2 vials
|
1 vial
|
4 ℃ / -20 ℃
|
|
Pre-Mixed Biotinylated Antibody (1:100)
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Enzyme Conjugate(1:100)
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Enzyme Diluent
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Antibody Diluent
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Standard Diluent
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Sample Diluent
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Washing Buffer (1:25)
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Color Reagent A
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Color Reagent B
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Color Reagent C
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Manual
|
1 set
|
1 set
|
RT
|
Place collected whole blood in refrigerator at 4 ℃ overnight. Then centrifuge for 10mins at 1000-3000 rpm. Take supernatant and either test immediately or place samples at -20 ℃/-80 ℃(1-3 months) for storage.
Take plasma where EDTA, sodium citrate, or heparin has been added as anticoagulant. Mix well. Centrifuge mixture for 10mins at 1000-3000 rpm. Take supernatant and test immediately or place samples at -20 ℃ /-80 ℃ (1-3 months) for storage.
Take tissue slices which have been washed in 0.01 M PBS and then a tissue protein extraction reagent has been added according to proportion of 1 g to 5-10 mL and then mixed on ice bath. After sufficient homogenization, please then centrifuge for 10mins at 5000-10000 rpm. Take supernatant for immediate testing, or place samples at -20 ℃/-80 ℃ (1-3 months) for storage.
Centrifuge for 10 mins at 1000-3000 rpm. Take supernatant for immediate testing, or place samples at -20 ℃/-80 ℃ (1-3 months) for storage.
Centrifuge for 10mins at 1000-3000 rpm. Take supernatant for immediate testing, or place samples at -20 ℃/-80 ℃ (1-3 months) for storage.
|
TYPICAL DATA
|
Reference STD1
|
Reference STD2
|
Reference STD3
|
Reference STD4
|
Reference STD5
|
Reference STD6
|
Reference STD7
|
Reference STD8
|
|
MMP-11, ng/mL
|
10
|
5
|
2.5
|
1.25
|
0.625
|
0.313
|
0.156
|
0
|
|
MMP-11, OD.
|
3.244
|
1.802
|
1.032
|
0.574
|
0.315
|
0.173
|
0.1
|
0.065
|
|
MMP-13, ng/mL
|
10
|
5
|
2.5
|
1.25
|
0.625
|
0.313
|
0.156
|
0
|
|
MMP-13, OD.
|
2.808
|
1.514
|
0.85
|
0.472
|
0.27
|
0.146
|
0.082
|
0.053
|
|
MMP-2, ng/mL
|
10
|
5
|
2.5
|
1.25
|
0.625
|
0.313
|
0.156
|
0
|
|
MMP-2, OD.
|
2.726
|
1.484
|
0.833
|
0.458
|
0.265
|
0.139
|
0.077
|
0.054
|
|
MMP-3, ng/mL
|
10
|
5
|
2.5
|
1.25
|
0.625
|
0.313
|
0.156
|
0
|
|
MMP-3, OD.
|
3.111
|
1.729
|
0.97
|
0.55
|
0.308
|
0.175
|
0.091
|
0.063
|
STANDARD CURVE
To assess linearity of the assay, samples were spiked with high concentrations of Rat MMP-11, MMP-13, MMP-2, MMP-3 and diluted with the appropriate calibrator diluent to produce samples with values within the dynamic range of the assay.
|
Matrix
|
Scope %
|
Average % Recovery Rate
|
|
serum(n=5)
|
82-94
|
88
|
|
EDTA plasma(n=5)
|
86-100
|
93
|
|
Cell Culture Media(n=5)
|
92-104
|
98
|
|
Within-Run Precision
|
Inter-Batch Precision
|
|||||
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
|
CV%
|
6.8
|
6.5
|
6.7
|
8.2
|
7.9
|
8.4
|