PART
|
96T
|
48T
|
STORAGE
|
|
Pre-Coated Plate
|
8×12
|
8×6
|
4 ℃ / -20 ℃
|
|
Pre-Mixed Standards
|
2 vials
|
1 vial
|
4 ℃ / -20 ℃
|
|
Pre-Mixed Biotinylated Antibody (1:100)
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Enzyme Conjugate(1:100)
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Enzyme Diluent
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Antibody Diluent
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Standard Diluent
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Sample Diluent
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Washing Buffer (1:25)
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Color Reagent A
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Color Reagent B
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Color Reagent C
|
1 vial
|
1 vial
|
4 ℃ / -20 ℃
|
|
Manual
|
1 set
|
1 set
|
RT
|
Place collected whole blood in refrigerator at 4 ℃ overnight. Then centrifuge for 10mins at 1000-3000 rpm. Take supernatant and either test immediately or place samples at -20 ℃/-80 ℃(1-3 months) for storage.
Take plasma where EDTA, sodium citrate, or heparin has been added as anticoagulant. Mix well. Centrifuge mixture for 10mins at 1000-3000 rpm. Take supernatant and test immediately or place samples at -20 ℃ /-80 ℃ (1-3 months) for storage.
Take tissue slices which have been washed in 0.01 M PBS and then a tissue protein extraction reagent has been added according to proportion of 1 g to 5-10 mL and then mixed on ice bath. After sufficient homogenization, please then centrifuge for 10mins at 5000-10000 rpm. Take supernatant for immediate testing, or place samples at -20 ℃/-80 ℃ (1-3 months) for storage.
Centrifuge for 10 mins at 1000-3000 rpm. Take supernatant for immediate testing, or place samples at -20 ℃/-80 ℃ (1-3 months) for storage.
Centrifuge for 10mins at 1000-3000 rpm. Take supernatant for immediate testing, or place samples at -20 ℃/-80 ℃ (1-3 months) for storage.
|
TYPICAL DATA
|
Reference STD1
|
Reference STD2
|
Reference STD3
|
Reference STD4
|
Reference STD5
|
Reference STD6
|
Reference STD7
|
Reference STD8
|
|
VCAM-1, ng/mL
|
50
|
25
|
12.5
|
6.25
|
3.125
|
1.563
|
0.781
|
0
|
|
VCAM-1, OD.
|
3.147
|
1.748
|
0.971
|
0.54
|
0.3
|
0.167
|
0.093
|
0.062
|
|
VE-cad, ng/mL
|
10
|
5
|
2.5
|
1.25
|
0.625
|
0.313
|
0.156
|
0
|
|
VE-cad, OD.
|
3.624
|
2.013
|
1.119
|
0.621
|
0.345
|
0.192
|
0.107
|
0.071
|
|
VLA-4, ng/mL
|
10
|
5
|
2.5
|
1.25
|
0.625
|
0.313
|
0.156
|
0
|
|
VLA-4, OD.
|
3.625
|
2.014
|
1.119
|
0.622
|
0.345
|
0.192
|
0.107
|
0.071
|
|
VCL, ng/mL
|
10
|
5
|
2.5
|
1.25
|
0.625
|
0.313
|
0.156
|
0
|
|
VCL, OD.
|
3.293
|
1.829
|
1.016
|
0.565
|
0.314
|
0.174
|
0.097
|
0.065
|
STANDARD CURVE
To assess linearity of the assay, samples were spiked with high concentrations of Human VCAM-1, VE-cad, VLA-4, VCL and diluted with the appropriate calibrator diluent to produce samples with values within the dynamic range of the assay.
|
Matrix
|
Scope %
|
Average % Recovery Rate
|
|
serum(n=5)
|
85-92
|
88.5
|
|
EDTA plasma(n=5)
|
92-99
|
95.5
|
|
Cell Culture Media(n=5)
|
95-107
|
101
|
|
Within-Run Precision
|
Inter-Batch Precision
|
|||||
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
|
CV%
|
6.4
|
6.3
|
6.5
|
8.5
|
8.2
|
8.5
|